Intraoral scan errors: seven to catch before you send the file

Reviewed by Dr. Vaishali Kesarwani, BDS, MBA (Healthcare Management) · UP State Dental Council Reg. 21804 · 23 Nov 2026

Illustration: A calm flat illustration of a handheld scanner wand beside a monitor showing an abstract arch mesh, with a simple tick-list card and soft te
AI-generated illustration

The research links intraoral scan errors mainly to wet or unretracted margins, poor lighting, mobile tissue, full-arch scanning and files that were never checked [1][2][3][4][5]. Most of these can be spotted at the chairside, before the scan reaches the lab.

The studies below measure accuracy. None of them counts redesigns, delays or remakes, so this article does not either. It lists seven errors where accuracy was shown to suffer, and a check for each.

1. Margins that are not exposed

A systematic review of crown preparations with subgingival margins found that scanner trueness was significantly compromised there. It often failed to reach clinically acceptable results [5].

The same review found that cord retraction improved scanner accuracy, and that acceptable results came with proper gingival retraction and a dry field [5].

Catch it: before export, rotate the scan and follow the margin all the way round. If any part is missing, blurred or filled in by the software, treat the scan as incomplete.

2. Saliva on the preparation

Saliva contamination reduced scanner accuracy in the same review, which also pointed to visibility problems [5]. A separate overview of systematic reviews lists saliva interference as one of the challenges of digital scanning [6].

Catch it: look for rough, pitted or distorted areas near the margin. They often mean moisture or poor visibility during capture. Dry the field and rescan that area.

3. Uncontrolled room lighting

A literature review reported that illuminance and colour temperature affected both the trueness and the precision of intraoral scanners [1]. Lighting is easy to overlook because it sits outside the scanner itself.

Catch it: note the operatory lighting during scanning. If a scan looks unusual, check whether the light changed, such as an overhead light being moved, and rescan under steady conditions.

4. Mobile and hard-to-trace tissue in edentulous arches

A systematic review of completely edentulous arches found that accuracy varied between scanners and between regions of the arch. Peripheral borders, the inner seal and poorly traceable structures such as the soft palate showed the largest discrepancies [2].

The same review found that clear anatomic landmarks, meaning hard tissue with attached mucosa, were recorded comparably to conventional impressions [2]. Its conclusion was that scanners can digitise denture-bearing areas, but the peripheral border and soft palate should be interpreted with care [2].

Catch it: do not assume the border is right because the surface looks smooth. Check it against what you saw in the mouth, and tell the lab which areas you are unsure about.

5. Full-arch implant scans taken as routine

A systematic review of full-arch digital implant impressions (30 studies, 29 of them in vitro) reported trueness from 7.6 to 731.7 μm. Using 100 μm and 0.4 degrees as acceptable limits, 18 studies reported deviations larger than 100 μm [4]. The authors concluded that full-arch digital implant impressions were not sufficiently accurate for clinical application [4].

Accuracy varied greatly with interimplant distance, scan body type, scanner type and operator experience [4]. The review found no effect from implant angulation, connection or depth [4].

Catch it: record which scanner and scan body were used. Ask the lab how it verifies full-arch implant files, for example with a verification jig or a conventional check. The sources do not say which verification method to use.

Summary of a systematic review of full-arch digital implant impressions: 30 studies, trueness 7.6 to 731.7 μm, 18 studies above 100 μm, factors that mattered and did not, and the authors' conclusion.
Full-arch implant impressions: what the systematic review reported.

6. Treating a scanner's range as unlimited

The literature review found that scanner repeatability suggested fixed prostheses could be made within the range of a partially edentulous arch [1]. It considered cross-arch fixed prostheses difficult with an intraoral scanner [1]. It also considered mouth guards and dentures possible with accuracy equivalent to desktop scanners [1].

For tooth-supported work, an overview of systematic reviews found digital scanning gave equivalent or better marginal adaptation than conventional impressions. Most of those reviews were rated low or critically low in quality [6]. That is a reason to be careful with the claim, not to drop it.

Catch it: match the scan to the case. For a long-span or cross-arch case, ask whether this scanner and workflow have been checked for that span [1].

7. Trusting the screen image

A review of scanner technologies warned that the rendering of files in the scanner's interface is often misleading. It discussed the accuracy of the files and of the recorded intermaxillary relationship as separate questions [3]. The same review lists scanning paths, tracking and mesh quality among the practical factors that affect a scan [3]. It also notes that powdering is a consideration with some systems [3].

The literature review adds that the accuracy of scanners must be verified [1].

Catch it: a clean-looking image is not proof of a sound file. Inspect the mesh for holes and stitching steps. Check that the bite registration matches how the teeth meet in the mouth.

What the evidence does not tell us

Most of the accuracy figures above come from in vitro studies or from reviews of mixed quality [4][6]. Accuracy also varies between scanners [2][4], so a finding for one system may not hold for another.

The sources do not say how often these errors lead to redesigns or remakes. They describe what reduces accuracy. They do not give rates of rework.

Checklist before you send a scan

Rapidontics Labs takes digital cases. Flagging doubtful areas in your notes helps whichever lab you use.

A checklist of eight checks to complete before sending an intraoral scan to the lab, from exposed margins to flagging doubtful areas.
Eight checks before a scan leaves the chairside.

Frequently asked questions

Does saliva affect intraoral scan accuracy?

In a systematic review of crown preparations with subgingival margins, saliva contamination reduced scanner accuracy. Clinically acceptable results came with proper gingival retraction and a dry field [5].

Can full-arch implant cases be scanned intraorally?

A systematic review concluded that full-arch digital implant impressions taken with intraoral scanners were not sufficiently accurate for clinical application. Accuracy varied with interimplant distance, scan body type, scanner type and operator experience [4]. Nearly all of its included studies were in vitro [4].

Which areas of an edentulous scan need extra care?

Peripheral borders, the inner seal and the soft palate showed the largest discrepancies. Hard tissue with attached mucosa was recorded comparably to conventional impressions [2].

Can I judge a scan from the screen image alone?

A review of scanner technologies says the rendering shown in the scanner's interface is often misleading. It treats the accuracy of the file and of the recorded bite relationship as separate questions [3]. The literature also says scanner accuracy must be verified [1].

Sources

  1. Accuracy and practicality of intraoral scanner in dentistry: A literature review.. Journal of prosthodontic research, 2020. https://doi.org/10.1016/j.jpor.2019.07.010
  2. Accuracy of Intraoral Scanner for Recording Completely Edentulous Arches-A Systematic Review.. Dentistry journal, 2023. https://doi.org/10.1186/s40510-015-0108-7
  3. Intraoral Scanner Technologies: A Review to Make a Successful Impression.. Journal of healthcare engineering, 2017. https://doi.org/10.1016/j.prosdent.2015.06.013
  4. Accuracy of full-arch digital implant impressions taken using intraoral scanners and related variables: A systematic review.. International journal of oral implantology (Berlin, Germany), 2021. https://pubmed.ncbi.nlm.nih.gov/34006079/
  5. Accuracy of Intra-Oral Scanners for Full Crown Tooth Preparations with Subgingival Margins: A Systematic Review.. The European journal of prosthodontics and restorative dentistry, 2025. https://doi.org/10.1922/EJPRD_2812Matias09
  6. Marginal adaptation of tooth-supported fixed restorations fabricated using digital scanning versus conventional impression techniques: An overview of systematic reviews.. The Journal of prosthetic dentistry, 2025. https://doi.org/10.1016/j.prosdent.2025.05.044

This article was drafted with AI assistance from the sources listed above and reviewed by Dr. Vaishali Kesarwani on 23 Nov 2026. It describes laboratory and digital workflows; clinical decisions stay with the treating dentist.